Laser Unequal Pathlength Interferometer
|
|
 |
| PRINT |
|
 |
| BACK |
|
The LUPI is a Twyman-Green type interferometer. In this type of instrument, light coming from the laser source is spatially filtered, collimated and passed through a beam splitter. Half of the light is reflected by the beam splitter and directed to a reference flat. The other half of the light passes through the beam splitter and is tightly focused. This focal point is also the focal point of an object being tested, such as a lens or a mirror.
After reflecting off the mirror being tested and reaching the reference flat, light is retro reflected. It retraces its path through the focal point and back to the beam splitter. In lens testing, beams collimated from the focal point also reach the reference flat and retrace their paths back to the beam splitter
At the beam splitter, the two beams of light recombine. Through constructive and destructive interference of the light waves, a pattern of light and dark fringes are formed. These are observed by the camera and fed to a video screen. The fringe pattern reveals the shape of the test piece and any errors that may be present. It can also be used to determine the focal length and off axis conditions.
The SORL LUPI Interferometer can easily be modified for phase shifting. For such an application a piezo operated flat can simply be placed and mounted in front of the permanent reference flat. The piezo induced phase shift in conjunction with the appropriate camera provides fringe analysis and interferogram interpretation.
List of all articles from this supplier
- OAP Series Off Axis Parabolic Mirrors
- Mirror Mounts for Parabolic and Flat Mirrors
- TOAN Series Off-Axis Reflective Blackbody Collimators
- COAR Series Laser Beam Expanders
List of all articles from this supplier
List of articles in this sub-category
- Newport Introduces the 1936-C and 2936-C Benchtop Optical Power and Energy Meters
- Iris Diaphragms
- Newport introduces 1918-C Handheld Optical Power and Energy Meter
- Safety Machine Windows
- Adaptors for coaxial incident light
- Modular directed lighting
List
of all articles in this sub-category Optical instrumentation, systems and accessories
List of all articles in this sub-categoryList of all articles in this sub-categoryList of all articles in this sub-category
|