vo-1 and vo-2 valve operators
|
|
 |
| PRINT |
|
 |
| BACK |
|
1. Extract plunger of operator VO-1 2. Place the adapter over valve and tighten. Finger tightening is sufficient. 3. Push plunger in to plug and screw in 2.5 to 3 turns. There is a mark on the knob. 4. Place the smooth end (0.375 inch/ 9.5mm O.D.) into vacuum system -- the connection may be adapter on pump or a high vacuum hose. 5. Rough pump manifold. Valve still closed.
6. Open valve. Pull out plunger and valve plug to open dewar valve. 7. Evacuate dewar to pressure less than 1E-6 Torr.
8. Close Valve. Push plunger in completely.
9. Close high vacuum valve and vent system, then unscrew plunger from valve plug. Then retract plunger.
10. Loosen the adapter and remove VO-1 from de war.
11. Replace dust cap on valve